| Model Number |
PCF-Q180AL |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the colonovideoscope for a separate issue.During the initial evaluation of the returned device, white residue in air water channel and cylinder.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the final investigation of the device that white residue inside the air water cylinder and channel observed.The initial evaluation findings were confirmed.Based on the results of the investigation, a definitive root cause could not be identified.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|