| Model Number |
106531US |
| Medical Device Problem Codes |
Failure to Charge (1085); Electrical Power Problem (2925)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient performed a system controller exchange at home after an emergency backup battery fault alarm.Upon review of the submitted log files, low power advisory events were observed that appeared to be associated with a poor connection between the battery contacts and the battery clip contacts.The log file recorded a backup battery fault event that appeared to have occurred after the system controller attempted a load test.A driveline disconnect event was also recorded.The alarm did not return after the battery was changed.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
|
| |
|