Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigation summary.The product was returned to eth for evaluation.Visual inspection revealed one labeled winding former with a needle suture combination outside the foil packet was received for analysis.Product code ep8766.During the initial inspection of the returned sample, a dented and broken suture was observed.The needle-suture combination was dispensed from the winding former and examined; no damage or defects were identified.It was further determined that the damaged suture fragment was not part of the needle-suture combination contained within the winding former.Therefore, the fragment was identified as foreign material.Based on the information currently available, a foreign matter was identified during the investigation of the sample received.This product issue will be addressed through eth quality system.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.
|