| Model Number |
M635WU60310 |
| Medical Device Problem Codes |
Migration (4003); Failure to Seal (4070)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that peri-device leak (pdl) occurred.The patient previously underwent a left atrial appendage (laa) closure procedure and a 31mm watchman flx pro closure device was implanted.At a follow up appointment on an unknown date, transesophageal echocardiogram (tee) visualized a pdl around the closure device and confirmed there was movement/shift of the device since the implant procedure.The patient subsequently returned for a leak closure procedure.The peri-device leak was successfully closed using two non-boston scientific (bsc) amplatzer vascular plug (avp) devices.The patient was reported to be stable with no additional complications.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete udi and other product specific information.If additional details become available a supplemental report will be submitted.
|
| |
|