| Catalog Number |
20027E |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that bd smartsite extension set was leaking.The following information was received by the initial reporter with the following verbatim.It was reported by customer that while preparing a csp and priming the line with drug the tubing was found to be leaking out of the 0.2-micron filter.The leaking tubing was identified prior to interaction with a patient.
|
| |
|
Additional Manufacturer Narrative
|
|
No product or photo was returned by the customer.The customer complaint of leakage from the fitler could not be verified due to the product not being returned for failure investigation.A device history record review was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.The root cause of this failure could not be identified without a sample investigation.The root cause in filter cases can be related to the medication used with the filter, causing the filter to wet-out of the air vents.
|
| |
|