This vmsr submission is part of a retrospective data remediation review.The included event represents one malfunction event that medline became aware of during q3 2024.Associated lot numbers: 65121110002, 65123040002, and 65123080001.Associated di: 10080196177813.No serial numbers were available.One malfunction event involving three devices is summarized in this report.Three devices were returned to the manufacturer and evaluated.The devices were not labeled for single use.No devices were identified as reprocessed or reused.The account reported that a medline hand-held aneroid sphygmomanometer, catalog number mds9388, would not inflate.Medline received three units in used condition and performed visual inspection and functional testing.Visual inspection found that the needle was away from the zero-off position on two devices.The needles were repositioned to the zero-off position using the screwdriver provided with the devices.Each device was then inflated to 300 mmhg, and testing was repeated five times per device.During testing, all cuffs inflated and deflated properly, the dials measured properly, and no leaks or other defects that would prevent the cuffs and gauges from working as intended were observed.The reported malfunction was not confirmed, and medline was unable to confirm that the product failed to meet specifications.A replacement was sent under reference 631083858.The supplier was not notified because the complaint was not confirmed with the returned samples.A retrospective review of the 12 months preceding the reported issue identified similar complaints involving mds9388 devices described as not inflating; however, no confirmed complaints were identified for lots 65121110002, 65123040002, or 65123080001, and no known trend was identified for these lots and this issue.The division will continue to monitor the issue for trending in accordance with sop-00032.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available, this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|