Model Number CYF-VH
Medical Device Problem Codes Corroded (1131); Crack (1135); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer returned the cysto-nephro videoscope for a separate issue.During the device analysis, the bending section cover glue was cracked off/detached, and the objective lens had corrosion.There was no patient involvement, and no harm was reported as a result of this event.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection.The reported issue was confirmed; the bending section cover glue was cracked off/detached, and the objective lens had corrosion could not be determined.However, the issue was likely a degradation problem and/or external factors.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Common Device Name
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50889286
Report Number3002808148-2026-33714
Device Sequence Number10663924
Product Code FAJ
UDI-Device Identifier04953170310461
UDI-Public04953170310461
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K221683
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCYF-VH
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-