|
The customer reported issue "leaking" with item mnk0077 (dressing,sheet,honey,therahoney,4x5).The issue was evaluated with received photos, and the supplier was notified of the issue.The reported issue was confirmed with the received photos.Root cause may be overfilling of the dressings and the supplier is investigating; the division will continue to monitor trending through the quarterly trend reports, and the complaint was ok to close.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|