The customer reported that the translator has fractured at the weld left side when sitting in it; fall did occur and some brusing but no medical attention was needed.Upon receiving the unit, the folding mechanism opened and closed properly, the right brake did not function properly and the right rear wheel rotated when the brake was engaged, the left brake did function properly, the backrest bar secured properly into both holsters, the seat upholstery was intact, and all four wheels were severely worn and dirty.Upon investigation, the right rear leg was fractured under the weld with the rear seat support bar, and the thickness of the rear leg at the fracture site measured 1.4mm, which is within specifications.The customer complaint has been confirmed; cause was cannot determine cause w/received sample, and no trends were identified for the specified material number and its corresponding defect type.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|