Catalog Number CR3434CTRH
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "bad cable broken pin.".
 
Additional Manufacturer Narrative
The reported issue was noted as "bad cable broken pin." the investigation included review of the original manufacturing records and evaluation of the device, including visual inspection; no observations were noted during the record review and the device passed all approved testing and inspections.Upon visual inspection, a bent pin was identified in the connector of the cable, and no other damage or defects were identified.The cause was identified as misuse/mishandling, traced to user, and medline renewal trends and monitors all complaints to ensure no systemic issues exist.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE CO-LABEL
Common Device Name
BW CATH CABL THERMCOOL 34PN BLK-34PN RED
Manufacturer (Section D)
Medline ReNewal
1500 ne hemlock avenue
redmond OR 94456 -187
Manufacturer Contact
brianna montgomery
three lakes drive
northfield, IL 60093
MDR Report Key50889568
Report Number3032391-2026-00177
Device Sequence Number9791558
Product Code IKD
UDI-Device Identifier10197344014163
UDI-Public10197344014163
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCR3434CTRH
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/26/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2023
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Patient SexUnknown
-
-