This complaint consists of tht registry data from japan.The information was obtained through the registry, no further details have become available for this event.It was reported that on (b)(6) 2026, a 25mm navitor vision valve was selected for implant using a small flexnav delivery system (ds).During the preparation, loading was not successful as while passing the valve through the ds, the tip of the valve does not expand uniformly, causing the struts to overlap.After multiple attempts, this situation remained unchanged so a new 25mm, navitor vision was replaced resulting in successful implantation without any complications.The patient was in a stable condition.It was reported through the tht registry from japan that on (b)(6) 2026, a 25mm navitor vision valve was implanted.On (b)(6) 2026, the patient experienced severe conduction disturbance and required a pacemaker implant.
|