• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC daVinci X; PATIENT SIDE CART, 4-ARM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC daVinci X; PATIENT SIDE CART, 4-ARM Back to Search Results
Model Number 380620-40
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
daVinci X
Common Device Name
PATIENT SIDE CART, 4-ARM
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50891430
Report Number2955842-2026-39895
Device Sequence Number17642682
Product Code NAY
UDI-Device Identifier00886874115404
UDI-Public(01)00886874115404
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K171294
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number380620-40
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/04/2026
Initial Report FDA Received Date09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/27/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-