It was reported that a non-preloaded monofocal intraocular lens (iol) was inserted into the patient's left eye and, during the implantation procedure, both haptics broke off inside the eye.The iol was subsequently removed and replaced with another lens of the same model and diopter power.No unplanned vitrectomy or additional medical intervention outside the standard of care was reported.It is unknown whether unplanned sutures were required or whether the procedure was delayed.The patient outcome was unknown.It was noted that the lens was discarded.No further information was provided.
|
|
Section a4 and a5: unknown, information was requested but not provided.Section d6a: if implanted, give date: not applicable, as lens was removed/replaced during the same surgery.Section d6b: if explanted, give date: not applicable, as lens was removed/replaced during the same surgery.Device evaluation:
product evaluation was not performed because the product was not received (the lens was discarded).The complaint issue reported could not be confirmed.Manufacturing record review:
the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search of complaints related to this production order (po) was performed.The search revealed that no other complaints were received for this po.Conclusion:
based on the investigation, no malfunction or product deficiency was identified, and the reported complaint issue could not be confirmed.Attempts have been made to obtain missing information; however, to date, the information has not been provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|