| Model Number |
PCF-H190DL |
| Medical Device Problem Codes |
Residue After Decontamination (2325); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the colonovideoscope displayed a scope communication error (e216) during inspection prior to use.During the initial evaluation of the device, white residue was observed within the auxiliary water channel.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The failure reported by the customer could not be confirmed; however, the failure identified during device inspection was confirmed.The root cause of the reported issue could not be determined.In addition, the root cause of the malfunction identified during device evaluation could not be established.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post market surveillance activities.
|
| |
|