It was reported that 2 hours and 38 minutes into dialysis treatment, the treatment was ended due to patient's drop in blood pressure and code team was called.Cardiopulmonary resuscitation (cpr) was initiated, despite medical intervention, the patient expired.The treating team attributed the event to the patient's preexisting conditions.No alleged tablo malfunction was reported or identified.
|
|
From the information provided, there is no indication that there was any device malfunction, nonconformance, or misuse that contributed to the reported event.Potential adverse events in the instructions for use (ifu) with the tablo system includes, but are not limited to, other, more serious, complications arising from dialysis, such as hemorrhage, air embolism, acidosis, alkalosis or hemolysis, can cause serious patient injury or death.Outset medical, inc.Has reviewed site device logs with a procedure date of (b)(6) 2026, no related system alarms were found to have occurred during treatment.Additionally, cross functional review of the treatment logs and event description confirmed normal device operation and did not identify any anomalous system activity that could have contributed to the reported event.Based on the available evidence, no device-related failure was identified.A review of production records for this serial number did not note any related manufacturing nonconformances that would contribute to this event.
|