| Catalog Number |
RSP0616MFSN |
| Medical Device Problem Code |
Incorrect Interpretation of Signal (1543)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that initially the spinplus nav 6x16mm 3 guide (b)(6) "was tracking accurately until midway through the case it stopped being accurate." inaccuracy was reportedly determined by "trying to get into frontal and showed the balloon in the brain.Inaccuracy was observed posterior.It was "off by 5 mm or more and then stopped tracking altogether".Staff opened a new spinplus which was accurate from then on.There was no patient injury or delay in surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Search Alerts/Recalls
|