It was reported that deflation failure occurred resulting in patient complications and additional intervention.The left anterior descending (lad) coronary artery was noted to provide two diagonal branches and terminated as the apical recurrent branch.The proximal lad was noted to have diffuse long 50 to 60% disease with 80 to 90% mid disease.Following atherectomy of the lad, direct stenting was performed with a 2.75 x 38 mm synergy xd drug eluting stent.The stent delivery system (sds) balloon was inflated for 10-15 seconds during stent implantation and post implant, was deflated for 20 seconds before attempting to pull back the device.The proximal segment remained inflated, however, and only the distal portion of the balloon was deflated.The sds was stuck in the patient for approximately 20 minutes and during that time, the patient reported chest pain and st elevations were noted.Multiple attempts were made to puncture the inflated balloon with different wires and micro catheters, but all failed.A 3-way stop cock was then connected and the technician continued to draw negative until the balloon deflated and was removed.The open-heart team had been called to perform further intervention to remove the sds, but the balloon was able to be deflated at the last moment before the patient was to be taken to the operating room.The balloon was inflated outside the patient to see if it was punctured or ruptured but it was able to maintain pressure.The patient condition post-procedure was stable and the patient was discharged.
|