| Model Number |
471172-19 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, a maryland bipolar forceps failed to perform its opening and closing function properly.It was not conducting energy.Upon inspection, it was observed that the wires were broken.The procedure was completed with no reported injury.Intuitive followed up and obtained additional information.The broken wires were the metallic wires controlling the movements of the instrument.No signs of thermal damage were observed at the site of breakage.No arcing was observed during the procedure.No photos were available since instrument was sent back.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the da vinci product with an alleged issue to perform failure analysis.However, failure analysis is not complete.
|
| |
|
Search Alerts/Recalls
|