Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The 8mm tip-up fenestrated grasper instrument was analyzed and found to was found to have the proximal clevis dislodged.As result of the dislodging, the main tube was broken.The dislodged proximal clevis is attached to the main tube.The grip cables are still intact and do not show damage.The main tube was broken at the distal end where the proximal clevis was attached.The component was broken and there were missing pieces, but the size of the missing pieces cannot be confirmed, scattered pieces.The complaint was confirmed by failure analysis.The probable root cause is attributed to damage during use.Applying excessive force on the distal end of the instrument during handling, insertion, usage (overloading), or removal can cause damage.
|