| Model Number |
380721-03 |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the customer reported that the bottom monopolar port on the integrated electrosurgical unit (iesu) was not functioning, displaying a question mark when instruments were attached.Different instruments and energy cords were tested but the port still failed to recognize them.The top monopolar port remained functional.The intuitive surgical, inc.(isi) technical support engineer (tse) log review showed no related errors.The procedure was completed, and no patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse went on-site and replaced the integrated electrosurgical unit (iesu).The system was verified and ready for use.The unit was analyzed and the bottom monopolar socket's pogo pin was jamm in the unit too far.The unit was connected to a golden system.All ports were tested and the bottom monopolar port was not working.
|
| |
|