| Model Number |
MR8-AA07 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that it was difficult to insert the tool into the attachment prior to use.At the time of report, there was no patient or staff impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: evaluation confirmed the reported malfunction of it being difficult to insert the tool into the attachment.The likely cause of the failure is due to the misaligned distal bearings of the attachment.It was also noted that the internal tube bearings were corroded.G3: aware date used as the date of analysis performed as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|