| Model Number |
ASY EMRGCY DRV UNT W/3 JNT HLD |
| Medical Device Problem Code |
Pumping Stopped (1503)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Livanova deutschland received a report that, while a patient was off the ecmo machine and undergoing a ct scan, centrifugal pump main unit¿s control panel suddenly lost power for 15 minutes, causing the centrifugal pump to stop.Normal operation was resumed after connecting the device to ac power.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A1-a5 patient information was not provided.H11
reportedly, the backup battery¿s charge level was inaccurately displayed; device battery had been replaced in august.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|