H3: product analysis: evaluation confirmed the reported malfunction of the attachment being blocked.The likely cause of the failure is due to the loose bearings at the distal tube of the attachment.It was also noted that the device was worn, it was not possible to insert the burr and test run the attachment, the bearings could not be inserted properly and was stuck and the tube markings were vague.G3: aware date used as the date of analysis performed as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|