| Model Number |
9560101 |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from multiple sources (manufacturer representative, healthcare professional, service repair) regarding a en doscope used for spinal therapy.It was reported that the instrument was not clear(cloudy). the screen went black and it improved when scope was replaced.Event occurred during intra-op and since event was discovered at an early stage, there was not much patient i nvolvement.There were no further complications reported regarding the event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis#: (b)(4): product: 9560101, lot no: 1943142 analysis confirmed that the reported issue (endoscope was not clear (cloudy)) was observed during the incoming inspection.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|