It was reported by the field service technician, that during routine check the cardiosave intra-aortic balloon pump (iabp) malfunctioned.Unit not getting on has been reported.No patient involved.No harm reported.The equipment was used on ac power and batteries.Fse arrived on site and check the machine, found machine power on normally without and error power input and basic functionally checked.Multiple power on/off cycles were performed and the cardosave machine operated normally after then perform the drive manifold test and pim test, all test passed, machine handed over to user in good working condition.
|