• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Alcon Laboratories Ireland Ltd. CLAREON PANOPTIX PRO Trifocal TORIC UV Absorbing IOL w/AutonoME Delivery System; Lens, Intraocular, Toric Optics

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Alcon Laboratories Ireland Ltd. CLAREON PANOPTIX PRO Trifocal TORIC UV Absorbing IOL w/AutonoME Delivery System; Lens, Intraocular, Toric Optics Back to Search Results
Model Number PXCAT3
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2026
Type of Reportable Event Malfunction
Event or Problem Description
A non-health care professional reported that during an intraocular lens (iol) implantation procedure, surgeon had an issue with two injectors on the same case.Both the devices were inoperable, so the surgeon manually loaded the second iol and completed the surgery.Additional information has been received and stated, the plunger was pushed past the lens during the lens insertion into the eye and in surgeon¿s opinion this contributed to or caused the event.The procedure was completed on the same day and the second lens still remains implanted.There was no harm or complication to the patient.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX PRO Trifocal TORIC UV Absorbing IOL w/AutonoME Delivery System
Common Device Name
Lens, Intraocular, Toric Optics
Manufacturer (Section D)
Alcon Laboratories Ireland Ltd.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
Alcon Laboratories Ireland Ltd.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
+1(800)757-9780
MDR Report Key50894799
Report Number9612169-2026-02186
Device Sequence Number10762358
Product Code MJP
UDI-Device Identifier00380658491482
UDI-Public(01)00380658491482(17)300817(11)250819(10)26088165
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPXCAT3
Device Catalogue NumberPXCAT3.280
Device Lot Number26088165
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/25/2026
Date Device Manufactured08/19/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Amvisc viscoelastic
Patient SexUnknown
-
-