Model Number SA60AT
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
A non-health care professional reported that during an intraocular lens (iol) implantation procedure, while loading the lens was found to be scratched/damaged.The procedure was completed on the same day.There was no contact between the patient eye and the device.The surgeon believed the lens was defective.Additional information has been requested but is not available at the time of this report.
 
Additional Manufacturer Narrative
The lens was returned positioned incorrectly in the lens case.Solution was dried on the lens.One haptic was folded and adhered to the anterior optic surface by solution.The optic had a deep scrape mark in the central optic zone on the posterior side of the lens.The photo provided appears to be of the returned sample.A batch record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the medical device file were met.Associated products were not provided.It is unknown if qualified products were used.The reported lens damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company¿s release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Associated products were not provided.It is unknown if qualified products were used.Company iols (intraocular lens) are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic visco surgical device (ovd) combination.The use of an unqualified combination may cause damage to iol and potential complications during the implantation process.The ifu (instruction for use) instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly, causing delivery issues and/or damage.The investigation has been completed based on current information.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaints are reviewed and monitored for adverse trends on a monthly basis.No adverse trends have been observed associated with the reported product and event.No further action has been identified for this reported event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action is warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AcrySof Singlepiece IOL
Common Device Name
Intraocular Lens
Manufacturer (Section D)
Alcon Research, LLC - Huntington
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
Alcon Research, LLC - Huntington
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
+1(800)757-9780
MDR Report Key50894918
Report Number1119421-2026-01943
Device Sequence Number6000975
Product Code HQL
UDI-Device Identifier00380652555821
UDI-Public(01)00380652555821(17)310531(11)260602(10)16269367
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSA60AT
Device Catalogue NumberSA60AT.210
Device Lot Number16269367
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/21/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/02/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-