| Model Number |
500004500000 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medbank tower, drawers were offline and the ns cabinet was stuck on the connecting page.The customer reported that the issue occurred when the user was trying to issue medications to a patient.However, there were no delay or adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 20-mar-2025 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, a field service engineer (fse) investigated the medbank and found no issues with the device.Further troubleshooting determined the problem was related to the hospital network.The hospital information technology team addressed and resolved the network issue, after which normal medbank functionality was restored.The system functioned as intended after the field service engineer investigated the device.
|
| |
|