• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION 303, INC. BD Pyxis¿ MedStation¿ ES Auxiliary; Automated Dispensing Cabinet

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION 303, INC. BD Pyxis¿ MedStation¿ ES Auxiliary; Automated Dispensing Cabinet Back to Search Results
Model Number 500001002502
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using the bd pyxis¿ medstation¿ es auxiliary system bottom panel paint had stripped and rusted.However, there were no delays or adverse events or injuries reported based on this event.
 
Additional Manufacturer Narrative
A review of the complaint history for sn (b)(6)was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6)was performed from the date of manufacture, 13-nov-2017 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of this incident, a field service engineer (fse) confirmed that the issue was resolved by the customer, and no further investigation was required.The system functioned as intended after the customer resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Pyxis¿ MedStation¿ ES Auxiliary
Common Device Name
Automated Dispensing Cabinet
Manufacturer (Section D)
CAREFUSION 303, INC.
3750 torrey view ct
san diego CA 92130
Manufacturer (Section G)
CareFusion 303, INC.
10020 pacific mesa blvd
san diego CA 92121 -438
Manufacturer Contact
brett wilko
3750 torrey view ct
san diego, CA 92130
+1(858)617-2000
MDR Report Key50895130
Report Number2016493-2026-66259
Device Sequence Number10712195
Product Code BRY
UDI-Device Identifier10885403512681
UDI-Public(01)10885403512681
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number500001002502
Device Catalogue Number156-033-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/13/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-