• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL Elektromedizinische Geraete GmbH SONATA; Cochlear Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL Elektromedizinische Geraete GmbH SONATA; Cochlear Implant Back to Search Results
Model Number Mi1260 SONATA 2
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Tissue Breakdown (2681)
Date of Event 09/21/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The user suffers from an implant extruding through the skin.As per new information received, the issue started on august 5th; implant telemetry showed the electrodes were functioning correctly, but calibration was not possible because the processors were broken.No skin damage was observed at that time.On august 25th, the mother reported the wound at the implant site.The mother did report to the audiologist that the child had become more aggressive since he stopped using the processors.The devices were sent for technical inspection, and a failure in the rcb was found in both units.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONATA
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL Elektromedizinische Geraete GmbH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43()577885626
MDR Report Key50895292
Report Number9710014-2026-00902
Device Sequence Number21503931
Product Code MCM
UDI-Device Identifier09008737385453
UDI-Public(01)09008737385453
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date08/10/2026
Device Model NumberMi1260 SONATA 2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2026
Initial Report FDA Received Date09/25/2026
Date Device Manufactured08/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age11 YR
Patient SexMale
-
-