| Model Number |
161001 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Hamilton medial ag received the following complaint description (summary):
it was reported that a hamilton-c1 (product no.161001, s/n: (b)(6), sw 3.0.3) experienced technical fault (tf) tf785003 (pm channel observation failed).Log review identified repeated occurrences of tf785003 together with tf444004 (voltage out of tolerance),
tf446029 (ambient failure detected), and tf485001 (ambient failure detected).No patient involvement, harm, or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Hamilton medical ag reference: (b)(4);
hamilton medical ags conclusion:
it was reported that a hamilton-c1 (product no.161001, s/n: (b)(6), sw 3.0.3) experienced technical fault (tf) tf785003 (pm channel observation failed).Investigation showed that no technical faults or technical events were recorded on the reported event date, 25-sep-2025.Review of the available logs identified repeated occurrences of tf785003 and tf444004 (voltage out of tolerance) on other dates.The investigation identified a voltage out-of-tolerance condition associated with the driver board as the most probable cause of the recorded technical faults.No patient involvement, harm, or medical intervention was reported.As correction, the driver board was replaced, and all applicable tests were completed successfully.Based on the investigation results, hamilton medical ag considers this case closed.
|
| |
|