It was reported that when using the bd pyxis¿ medbank tower drawers 9 and 10 were not working.As per part investigation, it was determined that the issue was identified as thermal damage to component u501 on the pcba pmc pyxis mini, resulting from an electrical failure.This damage compromised communication and control signals, causing the drawers to malfunction.However, there were no delays or adverse events or injuries reported based on this event.
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 29-nov-2018 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Part analysis:
the reported condition of drawer 9 and 10 not working was confirmed by fse and subsequently confirmed in the dchu inspection process.According to work order (b)(4), the field service engineer (fse) reported that replaced the module controller for drawers 9 and 10 due to no indicator lights, isolated and reconnected auxiliary cabinet sections for troubleshooting, coordinated medbank support reconfiguration and testing with no further failures observed, replaced faulty locks after cutting inaccessible rear locks, replaced the aib and internal pyxibus cable, migrated the all in one and bio id to the auxiliary unit for temporary main operation, unplugged drawer 14 per medbank support request, and confirmed all remaining drawers were functioning before monitoring continued.The system functioned as intended after the field service engineer repaired the device.During dchu visual inspection: o pn 151730-21: the inspection revealed localized thermal damage on component u501 (ic), characterized by a small burn mark and localized surface discoloration on the package.During dchu testing: o pn 151730-21: no functional testing was performed due to the observed thermal damage on component u501 on the returned module controller.The device was in use for treatment purposes as intended per 21 cfr 820.198(d).Root cause:
the root cause was identified as thermal damage to component u501 on the pcba pmc pyxis mini resulting from an electrical failure.This damage compromised communication and control signals causing the drawers to malfunction.
|