|
Device 26 of 50
based on the available information, this event is deemed to be a reportable malfunction.Photograph, video and/or physical sample evaluation:
photographs were received in association with this complaint and were evaluated as part of the investigation.Based on the review performed, the reported defect was confirmed from the evidence provided.Batch record revision results:
defect identified in the finished good manufacturing line:
lot 6c01271 was manufactured on 06/mar/2026 in convex 2 pc, with a total quantity of (b)(4) market units (mkus).A batch record review was performed by complaints engineer i on 14/sep/2026 for 1156500 and manufacturing order 1893970, as applicable.The review included verification of the applicable manufacturing procedures, bill of materials, component usage, tooling information, equipment settings, in-process controls, quality inspections, and final product release requirements relevant to the reported condition.Based on the documentation reviewed, the applicable procedures were followed, the required manufacturing and quality checks were completed, and the established acceptance criteria were met.No discrepancies related to the reported issue were identified within the batch record documentation reviewed.Defect potentially associated with the sub-assembly manufacturing line:
based on the nature of the reported defect, the condition could potentially occur during the sub-assembly process performed in the cvx sub manufacturing line.Therefore, a detailed batch record review was also performed for the applicable sub-assembly lot to determine whether any manufacturing or process-related discrepancy was identified that could have contributed to the reported condition.Sub-assembly lot 6c01272 was manufactured on 06/mar/2026 in cvx sub, with a total quantity of (b)(4) units.A batch record review was performed by complaints engineer i on 14/sep/2026 for 1156497 and manufacturing order 1893972, as applicable.The review included verification of the applicable sub-assembly manufacturing procedures, bill of materials, material usage, equipment settings, in-process controls, quality inspections, and applicable acceptance criteria relevant to the reported condition.Based on the documentation reviewed, the applicable procedures were followed, the required manufacturing and quality checks were completed, and the established acceptance criteria were met.No discrepancies related to the reported issue were identified within the finished good or sub-assembly batch record documentation reviewed.Historical complaints review:
on 14/sep/2026, complaints engineer i conducted a review of complaint records in database to identify complaints associated with lot number 6c01271 related to malfunction "skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (moldable only)".The review did not identify any additional complaints associated with the referenced lot.Historical nonconformance / capa review:
on 14/sep/2026, complaints engineer i performed a query in database to identify any in-process nonconformance(s) and/or corrective and preventive actions (capa) (s) associated with lots of numbers 6c01271 & 6c01272, under malfunction "skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (moldable only)".Based on the query results, no nonconformance(s) and/or corrective and preventive actions (capa) (s) related to this malfunction were generated during the manufacturing process of the referenced lot.Current quality controls:
based on process instructions, the following test method(s) are performed in the manufacturing line to detect this failure mode during the manufacturing process:
"srp centering on convex disc":
¿frequency: hourly
¿sample quantity: 5 samples
¿acceptance criteria: nlt 0.040" (1.0mm) from the outer diameter of the thermoformed disc to the inner diameter of the srp kiss cut collar."srp centering":
¿frequency: hourly
¿sample quantity: 5 samples
¿acceptance criteria: nlt 0.504" (12.8mm) the width of the srp on the left at any point in the frame."adhesive for centering convex discs (visual)":
¿ frequency: hourly
¿ sample quantity: 5 samples
¿ acceptance criteria: nlt 0.095" (2.4mm) from id of the adhesive center hole to id of the convex disc at any point.(qc tool 1289)
defect rate analysis:
as part of the investigation, a defect rate analysis was performed to evaluate the occurrence of the reported failure mode within the referenced lot.The calculation includes all confirmed defective units associated with the referenced lot identified at the time this investigation was performed, including the reported complaint and any additional complaint records identified during the historical complaint review.A total of (b)(4) defective unit(s) has been confirmed to date from a lot size of (b)(4) unit(s).This represents a defect rate of (b)(4)%.The calculated defect rate was compared against the applicable acceptance quality limit (aql) of 2.5 percentage, as defined in working instructions.Based on this comparison, the observed defect rate is below the established acceptable quality level (aql) threshold for this failure mode.Conclusion:
a review of the applicable batch record(s), quality records, and manufacturing documentation was completed as part of this investigation.The evaluation confirmed that all required manufacturing and quality activities were performed in accordance with established procedures and met the applicable acceptance criteria.No evidence was identified to indicate that the reported failure mode was associated with a manufacturing process deviation, documentation discrepancy, or quality system issue.This investigation will be closed.This issue will be monitored through the post market product monitoring review process, standard operating procedure (sop).To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number
reporting site: 1049092
manufacturing site: 9618003.
|