Following the reported event, user facility personnel discarded the assurance clip, and it is not available for return or evaluation.Without the return or evaluation of the device, a cause of the reported event could not be determined.Steris offered in-service training on the proper use and operation for the assurance clip however, the user facility declined.The instructions for use states "the device is suitable for endoscopes with a minimum channel diameter of 2.8mm.If the device is used in a non-forward viewing endoscope, the device may malfunction, or it may not be possible to detach the clip from the catheter.The passage of the clip through a retroflexed or tortuous path may result in the clip separating from the device system and potentially kinking or damaging the device.If the delivery system or clip part kinks or becomes damaged during device insertion or passage, do not use it."
no additional issues have been reported.
|