On-site investigation was performed by our field service engineer.The claimed issue was reproduced during troubleshooting.According to received information, the replacement of the support arm parts will solve the issue.The support arm serves to relieve the patient from the weight of the tubing system.It is worth mentioning that customer is obliged to follow the installation instructions delivered together with the device.For example, according to installation instructions for the support arm 177 (e.G.Rev.04) in the specification of the maximum capacity, it is stated how many kilograms can be loaded depending on the angle of use of the accessories (max.Approx.3 kg).When moving the support arm or changing position, the patient connection should be observed to see that no pulling or other movement occurs.The cause of the claimed issue in this customer product complaint has been determined.The issue was related to defective joints of support arm.Event site postal code: (b)(6).
|