| Model Number |
N/A |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Clinical Endoleak (4843)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Event reported from clinical study extend - nct05639400.Type 1c endoleak, was noted intraoperatively from right renal bridging stent on (b)(6) 2026 for patient (b)(6).Treatment of the event included a vbx stent placement on (b)(6) 2026.The event was considered resolved with treatment and sequalae on (b)(6) 2026.The patient continued in the study.The investigator considered the event of type 1c endoleak as serious with moderate severity and related to the device.The site provided an update on 28 aug 26 stating that this event is related to the thoraflex hybrid procedure and related to the thoraflex hybrid device.
|
| |
|
Additional Manufacturer Narrative
|
|
Clinical code:
4843 - clinical endoleak: site reported the event as a type1c endoleak.Impact code:
4624 - surgical intervention: the site reported that treatment of the event included a vbx stent placement on (b)(6) 2026.The event was considered resolved with treatment and sequalae on 27 july 26.The patient continued in the study.4642 - additional device required: site reported that treatment of the event included a vbx stent placement on (b)(6) 2026.Medical device code:
3190 - insufficient device problem information: awaiting further information from the site.Component code:
4755 - part/ component/ sub-assembly term not applicable.Type of investigation:
4109 - historical data analysis: four-year review was performed for thoraflex hybrid > leakage > endoleak > type 1, this gave an occurrence score of 0.132%.No trend was identified requiring action at this time.3331 - analysis of production records: comprehensive batch review was performed; no issue was identified.Device was manufactured to specification.4112 - analysis of information provided by user/third part: additional information was requested, awaiting further information from the site.Investigation findings:
3233 - results pending completion of investigation.Investigation conclusion:
11 - conclusion not yet available.22 - known inherent risk of device: review of ifu ( instruction for use) 30-213-usen rev 02 thoraflex hybrid us ifu section table 2 mentions endoleaks as potential adverse event.
|
| |
|
Search Alerts/Recalls
|