It was reported on an unknown date that they were writing to report an issue with the aiming arm they have in one of our four tibia nail advanced sets.They have used the aiming arm on the three separate occasions with issues using both the hybrid insertion handle and the stainless steel insertion handle.On all three occasions that they are reporting, it has missed the proximal static hole when they go to drill, which is causing a great deal of frustration with our surgeons.They are unsure of what else could be causing this issue at this point as we have also replaced the steel insertion handle, and all other sets seem to be okay.
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|