It was reported that during a cardiac ablation procedure, during pulsed-field delivery of the first application after creation of a post map, the sphere 9 catheter displayed a hashed appearance with bright electrodes, the green impedance graph was maximized, and the catheter was not displayed properly on the affera system.After disconnecting and reconnecting the sphere 9 catheter, it was nolonger visualized on the affera system.Troubleshooting included rebooting the affera system, magnet and then replaced cables, sphere 9 catheter and patches, moved the patient table and imaging equipment, but the issue persisted.The procedure was aborted and not continued using another system.An engineer tested the system the following day and was unable to reproduce the lack-of-visualization issue; the system functioned properly during testing.No patient complications have been reported as a result of this event.2026-09-25 it was later reported that the case was aborted under general anesthesia.
|
|
Continuation of d10: product id: afr-00016, product type: mapping system.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|