| Catalog Number |
05060303001 |
| Medical Device Problem Code |
Inaccurate Information (4051)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of a software issue with the accu-chek inform ii meter.The customer alleged that the formatting of the critical range displayed on the meter could cause a misinterpretation of the result.The customer alleged that the ranges listed under "ranges" are not the critical and/or reportable ranges but rather the acceptable ranges.Values that actually fall outside the displayed ranges would get flagged.The customer indicates that the ranges, as they currently display, are incorrect.There was no allegation of any actual misinterpretation of a result or of any patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Search Alerts/Recalls
|