The customer reported a quality concern of potential growth of mold on the chloraprep applicator, and the complaint was documented as dirty/stained chloraprep.Two photos were provided for the investigation, finished good production records were reviewed, and the finished good pack build was reviewed; no similar nonconformances related to the component were noted and no bom issues were identified.Two photos were provided of the chloraprep applicator with blue staining, confirming the reported issue.The investigation was limited due to the absence of a returned complaint sample, and it could not be determined whether the chloraprep applicator was activated prior to placement within the kit or occurred during transit or storage; as a result, a definitive root cause could not be determined at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|