• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: St. Jude Medical - Neuromodulation (Puerto Rico, LLC) Proclaim Plus Implantable Pulse Generator; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

St. Jude Medical - Neuromodulation (Puerto Rico, LLC) Proclaim Plus Implantable Pulse Generator; SCS IPG Back to Search Results
Model Number 3670
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 09/16/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the patient experienced inadequate stimulation with their scs system, pain at the ipg site and shocks when turning up the stimulation.As a result, surgical intervention was undertaken wherein the entire system was explanted.Note: it is unknown which lead is related to this issue.
 
Additional Manufacturer Narrative
Date of event is estimated.Additional components potentially involved in the event include: common device name: octrode, model: 3186, udi: (b)(4), serial: (b)(6), batch: a000167877.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Proclaim Plus Implantable Pulse Generator
Common Device Name
SCS IPG
Manufacturer (Section D)
St. Jude Medical - Neuromodulation (Puerto Rico, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
St. Jude Medical - Neuromodulation (Puerto Rico, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
sara speyerer
6901 preston road
plano, TX 75024
+1(972)309-8000
MDR Report Key50895948
Report Number3006705815-2026-06045
Device Sequence Number14952591
Product Code LGW
UDI-Device Identifier05415067046383
UDI-Public(01)05415067046383(10)A000170233(17)270221
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number3670
Device Catalogue Number3670
Device Lot NumberA000170233
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS ANCHOR.; SCS ANCHOR.; SCS LEAD.
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient Weight200 LB
-
-