(b)(4).Date sent 9/25/2026.D4 batch # a7000l.Investigation summary.The product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned with the distal tip of the shaft broken.In addition, the tyvek packaging was returned along with the instrument.In an attempt to replicate the reported incident, functional evaluation revealed that clips did not advance into the jaws upon the next trigger actuation.To assess the condition of the internal components, the device was disassembled.Internal examination identified a fractured feed link, which prevented proper clip advancement and was determined to be the cause of the feeding malfunction.Additionally, fifteen (15) clips were present within the clip track.No other anomalies or damage to the internal components were observed during the evaluation.Possible causes for the damage found may be inadvertent pressure placed on the device shaft through bending, pressing against the trocar, other devices on the back table, using the device as a retractor or moving heavy tissue with the device shaft.The event reported was confirmed and it is related to improper use of the device.Possible causes for the damage found is due to the jaws may have been restricted during firing, or not fully through the trocar during firing.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch a7000l number, and no non-conformances were identified.
|