| Catalog Number |
90911 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was alleged that the cable connector was found damaged during surgery setup.The exact cause of the damage could not be determined, as no specific event leading to the failure was identified.However, this type of connector damage is typically associated with excessive or misaligned force during connection or disconnection and is therefore considered most likely to be handling-related.As spare camera head was used and the procedure was successfully complete with the use of versius surgical system.The camera cable was subsequently replaced.No harm was reported in relation to this event.At the time of this report, no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Cmr surgical limited is submitting this report to comply with fda regulation 803.The device was inspected by the distributor.Mechanical damage was found, which was probably due to improper handling during connection and removal of the cable.Cmr surgical ltd does not consider this report and content to be an admission that its product is defective or that it has caused a death or serious injury.
|
| |
|