According to the customer the staples were "breaking in half" during use; the customer reported there were "no issues related to this concern" in regards to the patient¿s status and no serious injury, medical intervention, or follow-up care related to the reported incident.No sample or photo was received for the investigation; finished good production records were reviewed with no line down situations, no run without approvals, deviations, or non-conformance reports, and the bom and picklist were reviewed with all correct components and quantities added according to specifications.At this time, medline was unable to confirm the reported defect, as no physical sample or detailed photographic evidence was provided, and the cause was documented as sample required to determine cause.There has been no change to the reportability decision associated with this complaint, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|