The customer reported that something is wrong with the clamp and that the clamp doesn¿t work properly so it is giving the patient too many fluids.Medline received 1 ea (b)(6) iv administration set; the device was observed for any manufacturing defects or cosmetic defects, and upon functional testing the roller clamp was slid closed to prevent any flow of fluid and observed for any leaks, then readjusted to allow free-flow of liquid.The device worked as intended and the complaint was not confirmed.No further action was taken because objectionable trends were not seen for this issue; division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|