BOSTON SCIENTIFIC CORPORATION EMBLEM? S-ICD; Implantable Cardioverter Defibrillator (Non-Crt) Back to Search Results
Model Number 3501
Medical Device Problem Codes Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 03/01/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had inappropriate shocks due to the oversensing of noise.The patent had a cough and bronchitis a at the time of one of the shocks.The sensing vector was programmed to the primary vector.Also, the high energy shock impedance was 111 ohms, which is high but within normal limits.Technical services discussed some testing and programming options.There were no additional adverse patient effects.The system remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM? S-ICD
Common Device Name
Implantable Cardioverter Defibrillator (Non-Crt)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
road 698, lot no. 12
dorado PR 00646 -260
PR   00646-2602
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50896269
Report Number2124215-2026-51127
Device Sequence Number12225569
Product Code LWS
UDI-Device Identifier00802526599200
UDI-Public(01)00802526599200(17)251130
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P110042
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2025
Device Model Number3501
Device Catalogue Number3501
Device Lot Number253663
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2026
Initial Report FDA Received Date09/25/2026
Date Device Manufactured12/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age76 YR
Patient SexFemale
-
-