• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION The Q-TRAK Subcutaneous Electrode; Implantable Cardioverter Defibrillator (Non-Crt)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION The Q-TRAK Subcutaneous Electrode; Implantable Cardioverter Defibrillator (Non-Crt) Back to Search Results
Model Number 3400
Medical Device Problem Codes Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 08/29/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had an inappropriate shock due to the oversensing of noise.The patent was sitting in a vibrating chair at the time of the shock.Technical services reviewed the device history and found another inappropriate shock due to noise.There were some untreated episodes due to the same noise.Technical services discussed some testing and programming options.There were no additional adverse patient effects.The system remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
The Q-TRAK Subcutaneous Electrode
Common Device Name
Implantable Cardioverter Defibrillator (Non-Crt)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
229 avenida fabricante
san clemente CA 92672
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50896270
Report Number2124215-2026-51126
Device Sequence Number5852986
Product Code LWS
UDI-Device Identifier00802526563201
UDI-Public(01)00802526563201(17)160901(10)A115361
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
P110042/S000
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/01/2016
Device Model Number3400
Device Catalogue Number3010
Device Lot NumberA115361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2026
Initial Report FDA Received Date09/25/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age46 YR
Patient SexFemale
-
-