Model Number 3401
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the system was being checked prior to the patient getting an mri.Upon interrogation, the programmer displayed a high impedance message.The low energy impedances were greater than 400 ohms.Also, the therapy was off.There was some programming done earlier and the auto setup feature had not been run.Now, the device could not be programmed to the mri mode due to the high impedances, and the likelihood that the auto set feature had not been completed.Technical services advised if the auto setup feature was run previously, and a high impedance is encountered, it can cause the therapy to turn off.The doctor elected to leave the therapy off and do the mri anyway.A review of the device downloaded data confirmed the impedances have been increasing since 2022.There were no adverse patient effects.The system remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM S-ICD
Common Device Name
Implantable Cardioverter Defibrillator (Non-Crt)
Manufacturer (Section D)
Oscor Inc
3816 desoto blvd.
palm harbor FL 34683
Manufacturer (Section G)
Oscor, Incorporated
3816 desoto blvd.
palm harbor FL 34683
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50896346
Report Number2124215-2026-51130
Device Sequence Number16142847
Product Code LWS
UDI-Device Identifier00802526585968
UDI-Public(01)00802526585968(17)181001
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P110042/S043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/01/2018
Device Model Number3401
Device Catalogue Number3401
Device Lot NumberA138459
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2026
Initial Report FDA Received Date09/25/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
-
-