| Model Number |
BF-1TH190 |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the bronchovideoscope to the service center for a separate issue.During the device analysis, the olympus service repair technician indicated that there was a flickering image.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided if available.The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: an electrical/electronic component problem identified, which is an expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|