| Catalog Number |
04745914001 |
| Medical Device Problem Codes |
Mechanical Problem (1384); High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received an allegation about discrepant sodium electrode results for 1 patient sample tested on a cobas 6000 c501 module.Initial result: 197 mmol/l.During the validation process, the customer noticed that the initial result was elevated, prompting a repeat of the sample.Repeat result: 138 mmol/l.The repeat result was considered correct.No questionable result was reported outside the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The sodium electrode lot number and expiration date were not provided.The field service engineer (fse) found that the aspiration tubing in the wash chamber was partially torn at the vacuum vessel, causing sporadic dripping from the needle into random cuvettes.He replaced the wash station tubing set.The fse performed test runs with successful results.He then verified that the instrument was performing within specifications.The investigation is ongoing.
|
| |
|